SINGLE USE ONLY. Discard both electrodes after single use.
- Inspect application area for imperfections.
- Clean application areas with alcohol prep and allow to dry (do not abrade skin while cleaning).
- Remove active electrode by peeling backing from adhesive tape.
- Apply active electrode over treatment area, making sure there are no wrinkles or creases. Slightly stretching the tape while applying decreases wrinkling and promotes better skin adhesion.
- Remove hydrogel dispersive electrode (not included with 6570 series) from its plastic backing and place 4-6 inches away from treatment area, making sure the entire area is in contact with skin.
- Attach lead wires according to instrument instructions.
- Apply ionic solution into active electrode fill ports according to protocol.
Suggested Maximum Current
| Model | Max. Current Rating | Model | Max. Current Rating | Model | Max. Current Rating |
| 6550L | 4 mA | 6570EM | 4 mA | 6580M | 3 mA |
| 6550S | 4 mA | 6570M | 3 mA | 6580S | 1.5 mA |
| 6560L | 4 mA | 6570S | 1.5mA | 6580L | 4 mA |
| 6560EM | 4 mA | 6570R | 4 mA | 6580R | 4 mA |
| 6560M | 3 mA | 6580L | 4 mA | 6580LC | 4 mA |
| 6560S | 1.5 mA | 6580EM | 4 mA | 6580SC | 4 mA |
| 6570L | 4 mA |