SUPPLIES:
>> One bag of sterile infusion solution (water or saline)
>> One 1890 Transducer Holder
- bracketed to IV pole
- attached to Spectrum case
- attached to face of patient unit
>> For each pressure being measured (pves, pabd, pura or pcav):
- One (1) Model # 1880 Interchangeable Pressure Transducer
- One (1) APIS tubing
- One (1) sterile 18 Gauge 1" or 2.5cm needle
- 20 cc syringe Luer lock syringe (optional)
* NOTE: It has been documented that there is mixing of urine with the fluid in the device connected to the bladder cather, whether that device is a tubing (APIS) or the transducer. In addition, devices left sitting moist (transducers, tubings, etc.) quickly become contaminated with environmental flora (e.g., pseudomonas, staph, etc). Therefore, it is recommended that these products be changed between patients. Refer to Application Note: Reuse of Products Labeled "for Single Use only" for additional information.
| PROCEDURE: 1. Connect the Model 1880 Interchangeable Pressure Transducer to the Transducer Interface Cable. To connect, align the arrows, and then insert one connector into the other. To disconnect gently pull apart. 2. Insert the transducer (with the vented caps in place) into the 1890 transducer holder. 3. Remove the protective caps from the Luer fittings on the transducer. 4. Remove the protective cap from the female end of the APIS and attach to the male Luer at the top of the transducer. 5. Remove the protective cap from the male end of the APIS tubing and attach a sterile 18G, 1"/2.5 cm disposable needle. Discard the APIS protective cap. 6. Close the two-way stopcock at the base of the 1880 transducer (1880s are packaged with the stopcock open). 7. Spike the needle into the injection port of the solution bag hanging for infusion during CMG. NOTE: Use of an 18 Gauge or larger needle usually facilitates fairly rapid filling of the tubing and transducer from a fluid bag. 8. Open the two-way stopcock to permit the flow of solution from the bag, into the APIS tubing and transducer. 9. As the solution approaches the 1880 transducer - Remove the transducer from its bracket and invert it so that the end with the 2-way stopcock is up - Gently tap the transducer with the open palm of your hand, to ensure that no air bubbles are in the chamber. - NOTE: Do not tap the model 1880 transducer with metal objects (e.g. hemostats) to purge air bubbles. 10. Closely inspect the transducer for presence of air bubbles. If bubbles are detected, repeat steps above. 11. After eliminating all bubbles, close the stopcock and replace the transducer into the holder. 12. Remove the needle from the solution bag injection port and either recap or remove (leaving tip of APIS uncovered). 13. Place the tip of the APIS, (with or without capped needle) behind the superior connection (as shown), at the level of Symphysis Pubis for balancing. 14. The transducer is ready for use. |
PROCEDURAL OPTION: |





